CONFigure [1] |2|3|4 Commands

The CONFigure commands are used on the specified
window/measurement to set:
• The expected power level being measured.
• The resolution of the window/measurement.
• The channel(s) on which the measurement is to be made.
The CONFigure commands do not make the power measurement after
setting the configuration. Use READ?, or alternatively use INITiate
followed by a FETCh? to make the measurement.
The CONFigure command also applies the following defaults to the
channel(s) which are in the specified window (the channel(s) in the
window are specified in the <source list> parameter):
N1911A/1912A P-Series Power Meters Programming Guide
Default Settings
INITiate:CONTinuous OFF
TRIGger:SOURce IMMediate
TRIGger:DELay:AUTO ON
SENSE:AVERage:COUNt:AUTO ON
SENSE:AVERage:STATe ON
MEASurement Commands
Description
Sets the power meter to make one trigger cycle
when INITiate is sent.
When TRIG:SOUR is set to BUS or HOLD, sets
the power meter to make the measurement
immediately a trigger is received.
Enables automatic delay before making the
measurement.
Enables automatic filter length selection.
Enables averaging.
2