TRIGger Subsystem
INITiate:CONTinuous:ALL <boolean>
Sets all trigger sequences to be continuously initiated.
If INITiate:CONTinuous:ALL is set to:
• ON, trigger sequences are set to be continuously initiated
• OFF, trigger sequences are not set to be continuously initiated
Syntax
Example
INIT:CONT:ALL ON
Reset Condition
On reset (*RST), this command is set to OFF.
On preset (SYSTem:PRESet) and instrument power- up, when entering local
mode, if TRIGger[:SEQuence[1]|2]:SOURce is set to INT[[1]|2] or EXT,
INITiate:CONTinuous is not changed. For other trigger sources,
INITiate:CONTinuous is set to ON.
652
INIT
:CONT
Space
:ALL
This command sets all trigger sequences
to be continuously initiated.