• 1 is returned when trace capture is enabled
• 0 is returned when trace capture is disabled
Query Example
Error Messages
• If a P- Series or E- Series E9320 sensor is not connected, error –241,
• If an E- Series E9320 sensor is connected and set to AVERage mode
• If source is set to ON ( for dual channel, at either measurement
N1911A/1912A P-Series Power Meters Programming Guide
TRAC1:STAT?
"Hardware missing" occurs.
rather than NORMal mode, error –221, "Settings conflict" occurs.
channel) when the N1920 sensor is connected in normal mode and
SENse:BUFFer:COUNt or SENse:FREQuency:STEP is more than 1,
error –221, "Settings conflict" occurs.