[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
This command sets the measurement mode for the E9320 and N1920
sensor
either to average or peak sensor mode.
Syntax
SENS
SENS2
Parameters
• If TRIG:SOUR is set to INT1, INT2 or EXT, it is set automatically to IMM.
• INIT:CONT is set automatically to ON.
• SENS:AVER2:STAT is set automatically to OFF.
• CALC:FEED is set automatically to "POW:AVG" for all CALC blocks using the specified channel in
410
:
1
Item
Description/Default
character_data
Defines the measurement mode:
1
When measurement mode is set to average:
their CALC:MATH:EXPR.
2
When measurement mode is set to NORMal:
SENS:CORR:DCYC:STAT is set automatically to OFF.
:
FUNC
DET
AVERage: sets the E9320 and P-Series
sensor to average only mode.
NORMal: sets the E9320 and P-Series
sensor to normal mode.
N1911A/1912A P-Series Power Meters Programming Guide
Space
character_data
?
Range of Values
1
AVERage
2
NORMal