3
CALCulate Subsystem
Parameters
Example
CALC3:FEED2 "POW:AVER ON
SWEEP2"
Reset Condition
On reset, data_handle is set to :POW:AVER.
166
Item
Description
string
The input measurement type to be fed to the specific input
on the CALC block:
•
PEAK: peak power
•
PTAV: peak to average
•
AVER: average
•
MIN: minimum power
Values may be followed by ON SWEEP[1]|2|3|4
where the numeric specifies the gate to be used for the
feed. For example: "POW:PEAK ON SWEEP2".
If ON SWEEP[1]|2|3|4 is not supplied, the gate used is
left unchanged.
A feed of "" (empty string) disables the CALC block and
switches off that display line.
This command selects the input for FEED2
of CALC block CALC3 to be average power,
using gate 2. The channel from which the
feed is taken is determined by
CALC:MATH:EXPR..
N1911A/1912A P-Series Power Meters Programming Guide
Range of Values
"POW:PEAK"
"POW:PTAV"
"POW:AVER"
"POW:MIN"