PSTatistic:CCDF:MARKer[1]|2:DATa?

This command is used to retrieve the power and probability values at the
current marker position at any trace the marker is positioned.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
Syntax
Example
Reset Condition
On reset, the marker 1 and marker 2 will be set back to their default
positions.
N1911A/1912A P-Series Power Meters Programming Guide
PST
:CCDF
PST:CCDF:MARK1:DAT?
:DAT
1
:MARK
2
This command returns the power and
probability values at marker 1.
PSTatistic Subsystem
?
9
315