[Sense[1]]|Sense2:Frequency[:Cw|:Fixed]:Step <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXEd]:STEP
<numeric_value>
This command sets the number of steps in average or peak frequency
sweep. It must be used in conjunction with external trigger.
The frequency sweep range will be equally divided by the frequency step.
If trace display is turned on, the measurement window will be restored to
single numeric or analog depends on the number of measurement channel.
Determine the Right Step to be Set
N O T E
Number of frequency step can be calculated using equation below:
Step = f
where,
Step = Number of frequency step
f
start
f
stop
Interval = Frequency step size
Example
When f
to
Step = f
418
– f
+ Interval
stop
start
Interval
= Frequency sweep's start point
= Frequency sweep's stop point
= 1 GHz and f
start
stop
– f
+ Interval
stop
start
Interval
= 5 GHz – 1 GHz + 0.5 GHz
0.5 GHz
= 9
= 5 GHz with given interval of 0.5 GHz, the Step should be set
N1911A/1912A P-Series Power Meters Programming Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents