[Sense[1]]|Sense2:Sweep[1]|2|3|4:Time <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
This command sets the length of the time- gated period (time- gate length)
for time- gated measurements for the P- Series and E9320 sensors which
are set to NORMal mode. To set an E9320 sensor to NORMal mode, refer to
the command
<character_data>"
Syntax
Parameters
Example
SENS2:SWE3:TIME 0.001
440
"[SENSe[1]]|SENSe2:DETector:FUNCtion
on page 410.
:
SENS
1
SENS2
Item
Description/Default
numeric_value
The length of the time gated period in seconds.
Units are resolved to 1 ns.
: TIME
SWE
1
2
3
4
DEF: the default value is 100 µs
This command sets the length to 0.001
seconds.
N1911A/1912A P-Series Power Meters Programming Guide
Space
numeric_value
DEF
?
Range of Values
0 to 1 second
DEF