2
MEASurement Commands
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RELative
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to single channel with
relative mode on. The relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.
Syntax
CONF
1
2
3
4
Parameters
Refer to
parameters in this command.
100
:POW
:SCAL
Space
expected_value
DEF
"Optional Parameters"
Item
Description/Default
expected_value
A numeric value for the expected power
(for the expected
level. The units of measurement are dBm
power level)
and W. The default units are defined by
UNIT:POWer.
resolution
A numeric value for the resolution. If
unspecified the current resolution setting
is used.
:AC
:REL
,
resolution
DEF
on page 90 for additional details on the
N1911A/1912A P-Series Power Meters Programming Guide
,
source list
Range of Values
sensor dependent
1
DEF
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF

Hide quick links: