Optional Parameters
CONFigure, FETCh?, READ? and MEASure? have the following three
optional parameters:
• An expected power value
• A resolution
• A source list
Expected Power Value
An <expected_value> parameter is only required if you are using an
E- Series power sensor or N8480 Series power sensor (excluding Option
CFT). It has no effect on P- Series power sensor, 8480 Series power sensor
or N8480 Series power sensor with Option CFT. The value entered
determines which of the power sensor's two ranges is used for the
measurement. If the current setting of the power sensor's range is no
longer valid for the new measurement, specifying the expected power
value decreases the time taken to obtain a result.
90
CONfigure1?
upper window/upper measurement
CONFigure3?
upper window/lower measurement
CONFigure2?
lower window/upper measurement
CONFigure4?
lower window/lower measurement
N1911A/1912A P-Series Power Meters Programming Guide