Agilent Technologies N1911A Programming Manual page 691

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Parameters
Item
Description/Default
source
Available trigger sources:
BUS: the trigger source is the group execute trigger <GET> bus command, a
*TRG common command or the TRIGGER:IMMediate SCPI command.
EXTernal: the trigger source is the trigger input in the back panel.
HOLD: triggering is suspended. The only way to trigger the power meter is
to use TRIGger:IMMediate.
IMMediate: the trigger system is always true. If INITiate:CONTinuous
is ON the power meter is continually triggering free (free run mode). If an
INITiate:IMMediate command is sent a measurement is triggered then
the power meter returns to the idle state.
INTernal: either INT1 (Channel A) or INT2 (Channel B).
The trigger source is set to IMMediate on instrument power-up and when entering local
N O T E
mode.
The MEASure and CONFigure commands automatically set the trigger source to
IMMediate.
The READ? or MEASure commands should not be used if the trigger source is set to BUS
or HOLD.
Example
TRIG:SOUR IMM
Reset Condition
On reset, the trigger source is set to IMMediate.
N1911A/1912A P-Series Power Meters Programming Guide
This command configures Channel A for
immediate triggering.
TRIGger Subsystem
Range of Values
BUS
EXTernal
HOLDIMMediate
INTernal[[1]|2]
663
14