Agilent Technologies N1911A Programming Manual page 549

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

–241
–310
–310
–310
–310
–310
–310
–310
–310
–321
–330
N1911A/1912A P-Series Power Meters Programming Guide
Hardware missing
The power meter is unable to execute the command because either
no power sensor is connected or it expects an E-Series or N8480 Series power
sensor and one is not connected.
System error;Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected is for use with CW signals
only.
System error;Ch A Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected to Channel A is for use
with CW signals only.
System error;Ch B Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected to Channel B is for use
with CW signals only.
System error;Sensor EEPROM Read Failed - critical data not
found or unreadable
This indicates a failure with your E-Series or N8480 Series power sensor. Refer to
your power sensor manual for details on returning it for repair.
System error;Sensor EEPROM Read Completed OK but optional
data block(s) not found or unreadable
This indicates a failure with your E-Series or N8480 Series power sensor power
sensor. Refer to your power sensor manual for details on returning it for repair.
System error;Sensor EEPROM Read Failed - unknown EEPROM
table format
This indicates a failure with your E-Series or N8480 Series power sensor power
sensor. Refer to your power sensor manual for details on returning it for repair.
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example,
Linearity, Temp - Comp (temperature compensation).
This indicates a failure with your E-Series or N8480 Series power sensor power
sensor. Refer to your power sensor manual for details on returning it for repair.
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In
this instance the power meter detects power sensors connected to
both it's front and rear channel inputs.
Out of memory
The power meter required more memory than was available to run
an internal operation.
Self-test Failed;
The -330, "Self-test Failed" errors indicate that you have a problem
with your power meter. Refer to Contacting Agilent
Technologies on page 119 for details of what to do with your faulty
power meter.