Output:recorder[1]|2:Limit:lower <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

OUTPut:RECorder[1]|2:LIMit:LOWer <numeric_value>
This command sets the minimum scaling value for the specified recorder
output. The units used are dependent on the units currently set for the
CALC block specified in OUTPut:RECorder[1]|2:FEED <data_handle>.
Syntax
OUTP
:REC
Parameters
Example
OUTP:REC:LIM:LOW –90
Reset Condition
On reset, the minimum scaling value is set to –150 dBm.
N1911A/1912A P-Series Power Meters Programming Guide
1
2
Item
Description/Default
numeric_value
A numeric value for the minimum scaling
value. The units used—dBm, W or %—are
dependent on the units currently set for the
CALC block specified in
OUTPut:RECorder[1]|2:FEED
<data_handle>.
:LIM
:LOW
This command sets the minimum scaling
value to –90.
OUTput Subsystem
Space
numeric_value
?
Range of Values
–150 to +230 dBm
1 aW to 100 XW
0 % to 999 %
8
295