iDEN
The iDEN set- up provides:
• Average power in one iDEN training and data pulse
• Peak- to- average one iDEN training and data pulse
• Average power in a 90ms iDEN frame
The measurement is started by detecting the iDEN training burst—for
example the burst emitted by a mobile—using the internal RF level trigger.
Time gating is used to measure the average power in the following 15 ms
(data pulse). Gate 1 is used to measure this data pulse. The 90 ms frame
is also captured to measure the average power in the entire frame. Gate 2
is used to measure the 90 ms frame.
The following table shows the power meter presets when <character_data>
is set to IDEN. Commands not listed are preset according to their DEFault
values (for further information refer to

Table 12-69 iDEN: Power Meter Presets

Command
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
N1911A/1912A P-Series Power Meters Programming Guide
Table 12-
Setting
+800.000 MHz
NORM
E9321A/25A: OFF
E9322A/26A: OFF
E9323A/27A: OFF
N1921/2A: OFF
µ
Gate 1: 0
s
Gates 2 - 4: 0
SYSTem Subsystem
26):
Comments
Frequency setting
Measurement mode
Sensor video bandwidth
Delay between trigger point and time
gated period.