MEMory Subsystem
Example
MEM:TABL:FREQ
200kHz,600kHz
Query
MEMory:TABLe:FREQuency?
The query returns a list of frequency points for the table currently
selected. The frequencies are returned in Hz.
Query Example
MEM:TABL:FREQ?
Error Messages
• If more than 80 frequencies are in the list, error –108, "Parameter not
• If the frequencies are not entered in ascending order, error –220,
• If a table has not been specified using the MEMory:TABLe:SELect
• If a frequency is sent which is outside of the allowed frequency range,
282
allowed" occurs.
"Parameter error;Frequency list must be in ascending order" occurs.
command, the data cannot be entered into the table and error –221,
"Settings conflict" occurs.
error –222, "Data out of range" occurs.
This command enters frequencies of 200
kHz and 600 kHz into the currently
selected table.
This command queries the frequency
points in the currently selected table.
N1911A/1912A P-Series Power Meters Programming Guide