Parameters
Example
SENS2:CORR:GAIN1
Reset Condition
On reset, CFACtor|GAIN1 is set to 100 %.
Query
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut]
[:MAGNitude]? [MIN|MAX]
The query returns the current gain correction setting or the values
associated with MIN and MAX.
N1911A/1912A P-Series Power Meters Programming Guide
Item
Description/Default
A numeric value.
numeric_value
(for CFACtor and
GAIN1)
1
For example, a gain of 60 % corresponds to a multiplier of 0.6 and a gain of 150 % corresponds to
a multiplier of 1.5.
DEF: the default value is 100 %
MIN: 1 %
MAX: 150 %
This command sets a gain correction of
100% for Channel B.
SENSe Subsystem
10
Range of Values
1
1 to 150 PCT
DEF
MIN
MAX
389