[SENSe[1]]|SENSe2:BUFFer:COUNt <numeric_value>
This command sets the buffer size for average or peak measurement. It
must be used in conjunction with external trigger.
It can only be set when frequency sweep is disabled (FREQ:STEP 0).
Otherwise, this parameter will be automatically overwritten by frequency
sweep step. If trace display is turned on, the measurement window will be
restored to single numeric or analog depends on the number of
measurement channel.
This command is only applicable when used with 8480, N8480, E4410, E9300, E9320 or
N O T E
N1920 power sensors (average or peak).
Syntax
SENS
SENS2
Parameters
N1911A/1912A P-Series Power Meters Programming Guide
:
1
Item
Description
numeric_value
A numeric value for buffer size.
:COUN
FREQ
SENSe Subsystem
numeric_value
Space
?
Range of Values
1 to 2048
10
381