INITiate[1]|2[:IMMediate]

This command sets the power meter in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power meter memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to
This command performs the same function as
N O T E
INITiate:[IMMediate]:SEQuence[1]|2.
Syntax
Example
INIT2:IMM
Error Messages
If the power meter is not in the idle state or INITiate:CONTinuous is ON,
error –213, "INIT ignored" occurs.
N1911A/1912A P-Series Power Meters Programming Guide
"FETCh[1]|2|3|4 Queries"
INIT
1
2
on page 110 for further details.
:IMM
This command places Channel B in the
wait for trigger state.
TRIGger Subsystem