Ddt <Arbitrary Block Program Data>|<String Program Data - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

17
IEEE 488.2 Command Reference
*DDT <arbitrary block program data>|<string program data>
The *DDT (Define Device Trigger) command determines the power meter's
response to a GET (Group Execute Trigger) message or *TRG common
command. This command effectively turns GET and *TRG into queries, with
the measured power being returned.
Syntax
Parameters
740
Space
*DDT
Type
arbitrary block program
data
string program data
1
The <action> field of the parameter may contain:
FETC?
FETC1?
FETC2? (N1912A only)
*TRG
TRIG1
TRIG2 (N1912A only)
2
The first digit after the # indicates the number of following digits. The following digits indicate
the length of the data.
arbitrary block program data
string program data
?
Description
The command which is
executed on a GET or
*TRG.
N1911A/1912A P-Series Power Meters Programming Guide
Range of Values
1,2
#nN<action>
1
"<action>"