Trigger[1]|2:Source Bus|External|Hold|Immediate|Internal[[1]|2] - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

14
TRIGger Subsystem
TRIGger[1]|2:SOURce
BUS|EXTernal|HOLD|IMMediate|INTernal[[1]|2]
This command configures the trigger system to respond to the specified
source. This command only selects the trigger source. Use the INITiate
command to place the power meter in the wait for trigger state.
• This command has been included for compatibility purposes. It has the same purpose as
N O T E
• If the source is set to INT, INT1 or INT2 when connecting the N1920 or E9320 sensor in
Syntax
662
TRIGger[:SEQuence[1]|2]:SOURce
BUS|EXTernal|HOLD|IMMediate|INTernal[[1]|2] which should be used in
preference.
average mode, errors occurs.
TRIG
:SOUR
1
2
Space
HOLD
N1911A/1912A P-Series Power Meters Programming Guide
BUS
EXT
IMM
INT
1