SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the rear panel trig out
for testing purposes.
This command overrides the OUTPut:TRIGger[:STATe] command.
N O T E
For example, if OUTPut:TRIGger[:STATe] is ON and the command
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets it to
OFF. However, the 10 MHz remains out the Trig out port.
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now routed to
the Trig out overriding the service command turning the 10 MHz to off.
If the command is set to:
• ON, the 10 MHz time base signal is sent to the rear panel trigger out
• OFF, the 10 MHz time base signal is disabled.
Syntax
Example
SERV:BIST:TBAS:STAT OFF
N1911A/1912A P-Series Power Meters Programming Guide
connector.
:BIST
SERV
:TBAS
:STAT
This command disables the signal.
SERVice Subsystem
Space
0|OFF
1|ON
?
15