Memory:table:frequency <Numeric_Value>{,<Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

MEMory:TABLe:FREQuency
<numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected
table. Any previous frequency list is cleared before the new frequency list
is stored. The frequencies must be entered in ascending order. Entries in
the frequency lists correspond as shown in
calibration/offset factor lists.
For sensor calibration tables only, the first calibration factor entered using the
N O T E
MEMory:TABLe:GAIN command is used as the reference calibration factor.
Predefined sensor calibration table is not applicable for N8480 Series power sensors with
N O T E
Option CFT. Therefore you are required to create a new sensor calibration table for the
sensors when a sensor calibration table is needed.
N1911A/1912A P-Series Power Meters Programming Guide
MEMory Subsystem
Table 7- 21
with entries in the
7
279