Agilent Technologies N1911A Programming Manual page 639

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Keyword
:TRANsition[1]|...|10
:REFerence?
:STATe
:UNIT
When making trace measurements, use the following command sequence to synchronize
N O T E
the returned trace data with the measurement:
N1911A/1912A P-Series Power Meters Programming Guide
:SEParation?
:NEGative
:DURation?
:OCCurrence?
:POSitive
:DURation?
:OCCurrence?
Command
TRIG:SOUR INT
or
TRIG:SOUR EXT
INIT:CONT OFF
1
TRAC:STAT ON
AVER:STAT OFF
or
TRIG:DEL:AUTO OFF
INIT
FETCH?
1
TRACE:DATA? MRES
1
A trace display format must be set when this command is used.
Parameter Form
<numeric_value>
<boolean>
<character_data>
Comment
Change trigger source to internal or external
Trace data can only be retrieved with
INIT:CONT OFF
Enables trace capture
No settling time delays for digital filter to fill
Initiates a new measurement
Fetch the result (waits for the measurement to complete)
Retrieves the trace data once the measurement has completed
TRACe Subsystem
Notes
Page
[query only]
page 626
[query only]
page 628
[query only]
page 630
[query only]
page 632
[query only]
page 634
[query only]
page 636
page 638
page 640
13
611

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents