17
IEEE 488.2 Command Reference
*ESR?
The *ESR? query returns the contents of the Standard Event Status
Register then clears it. The format of the return is <NR1> in the range of
0 to 255.

Table 17-91*ESR? Mapping

Syntax
744
Table 17- 91
Bit
Weight
Meaning
0
1
Operation Complete
1
2
Request Control (not used)
2
4
Query Error
3
8
Device Dependent Error
4
16
Execution Error
5
32
Command Error
6
64
Not used
7
128
Power On
?
*ESR
shows the contents of this register.