Agilent Technologies N1911A Programming Manual page 498

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

11
STATus Subsystem
Parameters
Query
:ENABle?
The query returns a 15 bit decimal- weighted number representing the
contents of the Enable Register of the SCPI Register Set being queried.
The format of the return is <NR1> in the range of 0 to 32767 (2
:NTRansition <NRf>|<non-decimal numeric>
This command sets the Negative Transition Register of the SCPI Register
Set you require to control. The parameter value, when rounded to an
integer and expressed in base 2 has its first 15 bits written into the
Negative Transition Register of the SCPI Register Set concerned. The last
bit (bit 15) is always set to 0.
470
Type
NRf
non-decimal numeric
Description
The value used to set the
Enable Register.
N1911A/1912A P-Series Power Meters Programming Guide
Range of Values
16
0 to 2
–1
15
–1).