System:communicate:lan:aip[:State] <Boolean - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

12
SYSTem Subsystem
SYSTem:COMMunicate:LAN:AIP[:STATe] <boolean>
This command enables the AutoIP protocol to dynamically assign the IP
address when connecting to the power meter in an isolated (non- site)
LAN network (for example, laptop to power meter).
Syntax
Example
SYST:COMM:LAN:AIP ON
Query
SYSTem:COMMunicate:LAN:AIP?
• 1 is returned if AutoIP is enabled
• 0 is returned if AutoIP is disabled
Query Example
SYST:COMM:LAN:AIP?
496
:COMM
SYST
:AIP
:LAN
This command enables the AutoIP
This command queries the state of the
AutoIP.
N1911A/1912A P-Series Power Meters Programming Guide
Space
:STAT
0|OFF
1|ON
?

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents