PSTatistic:CCDF:GAUSsian:MARKer[1]|2[:SET]
This command is used to set the markers on Gaussian trace. The markers
will be set only if the trace is present and visible. According to the
selections made, the markers will become active on the screen.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
• Expanded window is enabled (not full CCDF screen window)
Syntax
Example
Reset Condition
On reset, the marker is set on the next visible trace according to the
sequence Channel A, Channel B, Reference or Gaussian.
N1911A/1912A P-Series Power Meters Programming Guide
PST
:CCDF
PST:CCDF:GAUS:MARK
:MARK
:GAUS
This command sets marker 1 on the
Gaussian trace.
PSTatistic Subsystem
1
:SET
2
9
311