TRACe[1]|2:MEASurement:PULSe[1]|...|10:DCYCle?

Algorithm
Syntax
Example
Error Messages
This command returns the duty cycle of the selected pulse in percentage.
Duty Cycle = (pulse duration / pulse period) * 100
where,
pulse duration is the time difference between positive and negative
transitions of one pulse and pulse period is the time difference between
two consecutive transition occurrences of the same polarity.
TRAC
1
2
TRAC2:MEAS:PULS3:DCYC?
• The command is only applicable when N192x or E932x Sensors are
:PULS
:MEAS
This command returns the duty cycle of
the 3rd pulse found on trace 2.
1
:DCYC
2|...|10
?