Trace[1]|2:Measurement:pulse[1]|...|10:Separation - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

13
TRACe Subsystem

TRACe[1]|2:MEASurement:PULSe[1]|...|10:SEParation?

This command returns the time difference of the n
found on a trace. As power pulses are by definition positive pulses, the
pulse separation is the time difference between negative transition of one
pulse and the positive transition of the next pulse.
Algorithm
If the first transition in the trace is positive,
then
PULSe:SEParation = time the second positive transition occurs - time the
first negative transition occurs
else
PULSe:SEParation = time the first positive transition occurs - time the first
negative transition occurs.
Syntax
Example
626
TRAC
1
2
TRAC1:MEAS:PULS:SEP?
:MEAS
:PULS
This command returns the time
separation of the 1st and 2nd pulses found
on trace 1.
N1911A/1912A P-Series Power Meters Programming Guide
th
and (n+1)
:SEP
1
2|...|10
th
pulses
?