Editing Frequency Dependent Offset Tables

It is not possible to create any additional frequency dependent offset
tables. However, the 10 existing tables can be edited using the MEMory
subsystem. To do this:
1 Select one of the existing tables using:
2 Enter the frequency data using:
3 Enter the offset factors as shown in the table below using:
4 If required, rename the frequency dependent offset table using:
N1911A/1912A P-Series Power Meters Programming Guide
MEMory:TABle:SELect <string>
For information on naming frequency dependent offset tables see
"Naming Frequency Dependent Offset Tables"
information on the current names which you can select refer to
the Frequency Dependent Offset Table Names"
MEMory:TABle:FREQuency <numeric_value> {,<numeric_value>}
MEMory:TABle:GAIN <numeric_value> {,<numeric_value>}
Frequency
Frequency 1
Frequency 2
"
Frequency n
MEMory:TABLe:MOVE <string>,<string>. The first <string> parameter
identifies the existing table name, and the second identifies the new
table name.
Power Meter Remote Operation
on page 31. For
on page 30.
Offset
Offset 1
Offset 2
"
Offset n
1
"Listing
29