Table 12-42 Bluetooth: Power Meter Presets - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

12
SYSTem Subsystem
BLUetooth
The Bluetooth set- up provides:
• Average power in a Bluetooth DH1 data burst.
• Peak power in the same burst
• Display of RF pulse in one timeslot
The measurement is started by detecting the Bluetooth RF burst using the
internal RF level trigger. The internal trigger is set to –20 dBm.
Time- gating is used to measure the peak and average power in a single
Bluetooth DHI data burst which lasts for 366 us. The DHI burst does not
occupy a full Bluetooth timeslot, which lasts for 625 µs.
The following table shows the power meter presets when <character_data>
is set to BLUetooth. Commands not listed are preset according to their
DEFault values (for further information refer to

Table 12-42 BLUetooth: Power Meter Presets

Command
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Trigger setup
TRIG[:SEQ[1]|2]:SOUR
INIT:CONT
552
Setting
+2400.000 MHz
NORM
E9321A/25A: DEF
E9322A/26A: OFF
E9323A/27A: OFF
N1921/2A: OFF
Gate 1: 0.2 µs
Gates 2 - 4: 0
Gate 1: 366 µs
Gates 2 - 4: 0
INT1
ON
N1911A/1912A P-Series Power Meters Programming Guide
Table 12-
26):
Comments
Frequency setting
Measurement mode
Sensor video bandwidth
Delay between trigger point and time
gated period.
Length of time gated period for time
gated measurements.
Trigger source set up and acquisition