Example
CONF1:POW:AC DEF,2,(@1)
N1911A/1912A P-Series Power Meters Programming Guide
Item
Description/Default
source list
The channel which the command is
implemented on.
If unspecified the current window setup is
used. However, on the N1912A, if the
window shows a ratio or difference
measurement, the upper window defaults to
Channel A and the lower window to Channel
B.
1
The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the
command sub-systems. The parameters must be entered in the specified order. If parameters are
omitted, they default from the right. The parameter DEFault is used as a place holder. Specifying
DEF leaves the parameter value unchanged.
2
When the measurement result is linear this parameter represents the number of significant
digits. When the measurement result is logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001
respectively.
3
N1912A only.
MEASurement Commands
This command configures the upper
window/upper measurement to measure
the power of Channel A, using the current
sensor range and a resolution setting of 2.
Range of Values
(@1)
3
(@2)
99
2