Agilent Technologies N1911A Programming Manual page 50

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

1
Power Meter Remote Operation
setting the filter length to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 INIT2
6 FETC1? -50,DEF,(@2)
Example 4 - Specifying the Resolution Parameter
The previous examples detailed the use of the expected value and source
list parameters. The resolution parameter is used to set the resolution of
the specified window. This parameter does not affect the resolution of the
data, however it does affect the auto averaging setting (refer to
on page
Since the filter length used for a channel with auto- averaging enabled is
dependent on the window resolution setting, a conflict arises when a given
channel is set up in both windows and the resolution settings are
different. In this case, the higher resolution setting is used to determine
the filter length.
The following example uses the resolution parameter to specify a
resolution setting of 3. This setting represents 3 significant digits if the
measurement suffix is W or %, and 0.01 dB if the suffix is dB or dBm (for
further details on the resolution parameter refer to the commands in
Chapter
expected power and source list parameters are defaulted. The expected
power value is left unchanged at its current setting. The source list
parameter is defaulted as described in the note
page 15. Note that as the source list parameter is the last specified
parameter you do not have to specify DEF.
22
35).
2, "MEASurement Commands"). Also, in this example the
N1911A/1912A P-Series Power Meters Programming Guide
Figure 1- 2
"Agilent N1912A Only"
on