7
MEMory Subsystem
MEMory:CLEar[:NAME] <character_data>
This command clears the contents of a specified sensor calibration table
(8480 Series sensors and N8480 Series sensors with Option CFT only),
frequency dependent offset table, or save/recall register.
Although the table remains, a MEMory:TABLe:FREQuency|GAIN:POINts?
query returns a 0 as there are no contents in the table.
For sensor calibration tables and frequency dependent offset tables, this
command is an alternative form of the MEMory:CLEar:TABLE command,
the only difference being the method in which the table is selected.
The contents cleared using this command are non-recoverable.
N O T E
266
N1911A/1912A P-Series Power Meters Programming Guide