Pstatistic:ccdf:gaussian[:State] <Boolean - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

PSTatistic:CCDF:GAUSsian[:STATe] <boolean>
This command is used to turn on or off the Gaussian trace and it is
independent of the channels attached.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
• Expanded window is enabled (not full CCDF screen window)
Syntax
Example
Reset Conditions
On reset, the Gaussian trace will be cleared (OFF).
N1911A/1912A P-Series Power Meters Programming Guide
:CCDF
PST
PST:CCDF:GAUS ON
PST:CCDF:GAUS OFF
:STAT
:GAUS
This command turns on the Gaussian
trace.
This command turns off the Gaussian
trace.
PSTatistic Subsystem
Space
0|OFF
1|ON
?
9
309