PSTatistic[1]|2:CCDF:TRACe:POWer:PTAVerage?

This command is used to retrieve peak to average power value of Channel
A or Channel B trace.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
Syntax
Example
PST:CCDF:TRAC:POW:PTAV?
Error Messages
• If no power sensor is connected, error –241 "Hardware missing" occurs.
• If sensor/sensors connected are not P- Series sensors, error –241
• If measurement speed setting is FAST, error –221 "Settings conflict"
• If command is executed in other window besides CCDF window, error
N1911A/1912A P-Series Power Meters Programming Guide
PST
1
2
"Hardware missing" occurs.
occurs.
–221 "Settings conflict: Requires CCDF window" occurs.
:CCDF
:TRAC
This command returns the peak to
average power value of Channel A trace.
PSTatistic Subsystem
:POW
:PTAV
9
?
355