OUTPut:TRIGger[:STATe] <boolean>
This command enables or disables the trigger output signal.
When sensor is in triggered average measurement mode, the trigger output
signal will only be asserted after the measurement has settled.
This command is also applicable when used with 8480, N8480, E4410, E9300 or E9320
N O T E
sensor (Average mode only).
Syntax
Example
OUTP:TRIG:STAT 1
Reset Condition
On reset, the trigger output signal is disabled.
N1911A/1912A P-Series Power Meters Programming Guide
OUTP
:TRIG
Space
:STAT
This command enables the trigger output
signal.
OUTput Subsystem
0|OFF
1|ON
?
8