TRIGger[:SEQuence[1]|2]:SOURce
BUS|EXTernal|HOLD|IMMediate|INTernal[[1]|2]
This command configures the trigger system to respond to the specified
source. This command only selects the trigger source. Use the INITiate
command to place the power meter in the wait for trigger state.
This command has the same purpose as TRIGger[1]|2:SOURce
N O T E
BUS|EXTernal|HOLD|IMMediate|INTernal[[1]|2].
Syntax
TRIG
N1911A/1912A P-Series Power Meters Programming Guide
:SEQ
1
2
Space
:SOUR
TRIGger Subsystem
BUS
EXT
HOLD
IMM
INT
1
2
?
683
14