Agilent Technologies N1911A Programming Manual page 89

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

The standard event register is cleared when you:
• send a *CLS (clear status) command.
• query the event register using the *ESR? (event status register)
The standard event enable register is cleared when you:
• cycle the instrument power.
• execute a *ESE 0 command.
Questionable Status Register
The questionable status register provides information about the quality of
the power meter's measurement results. Any or all of these conditions can
be reported in the questionable data summary bit through the enable
register. You must write a value using the STATus:QUEStionable:ENABle
command to set the enable register mask.
The questionable status model is shown in the pullout at the end of this
chapter.
The following bits in these registers are used by the power meter.
N1911A/1912A P-Series Power Meters Programming Guide
Bit
Decimal
Number
Value
3
8
4
16
5
32
6
64
7
128
command.
Power Meter Remote Operation
Definition
Device Error
A device error occurred, refer to error numbers 310 to 350 in the
user's guide.
Execution Error
An execution error occurred, refer to error numbers 211 to 241
in the user's guide.
Command Error
A command syntax error occurred, refer to error numbers 101 to
161 in the user's guide.
User request.
Power On
Power has been turned off and on since the last time the event
register was read or cleared.
1
61

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents