Fetch[1]|2|3|4[:Scalar][:Power:ac]:Difference? [<Expected_Value>[,<Resolution>[,<Source List>]]] - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
MEASurement Commands
FETCh[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to power
difference with relative mode off, recalculates the measurement and places
the results on the bus. The result is a power based measurement and is
expressed in the units defined by UNIT[1]|2|3|4:POWer.
Syntax
FETC
1
2
3
4
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected
power level)
116
:POW
:SCAL
expected_value
Space
DEF
"Optional Parameters"
Description/Default
The expected power level parameter can be set to
DEF or a numeric value. If a value is entered it
should correspond to that set by CONFigure
otherwise an error occurs.The units of
measurement are dBm and W. The default units are
defined by UNIT:POWer.
:DIFF
:AC
,
resolution
DEF
on page 90 for additional details on the
N1911A/1912A P-Series Power Meters Programming Guide
?
,
source list
Range of Values
sensor dependent
1
DEF