1
Power Meter Remote Operation

Using Device Clear to Halt Measurements

Device clear is an IEEE- 488 low- level bus message which can be used to
halt measurements in progress. Different programming languages and
IEEE- 488 interface cards provide access to this capability through their
own unique commands. The status registers, the error queue, and all
configuration states are left unchanged when a device clear message is
received. Device clear performs the following actions.
• All measurements in progress are aborted.
• The power meter returns to the trigger "idle state".
• The power meter's input and output buffers are cleared.
• The power meter is prepared to accept a new command string.
For interfaces the that do not support a low-level device clear, use the ABORt command.