Example
PST1:CCDF:PROB? 50
Error Messages
• If no power sensor is connected, error –241 "Hardware missing" occurs.
• If sensor/sensors connected are not P- Series sensors, error –241
• If measurement speed setting is FAST, error –221 "Settings conflict"
• If the acquisition mode is in continuous triggering or triggering source
• If the parameter specified is less than 0.0 or more than 50.0, error –220
• If no parameter is specified, error –109 "Missing parameter" occurs.
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
occurs.
is set to either INT1, INT2, or EXT, error –221 "Setting conflict"
occurs.
"Parameter error" occurs.
This command queries the probability at
the power level of 50dB for Channel A.
PSTatistic Subsystem
9