Syntax
Query Example
CAL1:ALL?
Error Messages
• If the calibration was not carried out successfully the error –231, "Data
• If zeroing was not carried out successfully the error –231, "Data
N1911A/1912A P-Series Power Meters Programming Guide
CAL
:ALL
1
2
Questionable; CAL ERROR" occurs. If you are using an N1912A the
error message specifies which channel failed calibration.
Questionable; ZERO ERROR" occurs. If you are using an N1912A the
error message specifies which channel failed calibration.
If there is no sensor connected, the error –241, "Hardware Missing"
occurs.
CALibration Subsystem
?
This command causes the power meter to
perform a calibration sequence on
Channel A and return a result.