Output:recorder[1]|2:Limit:upper <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

OUTPut:RECorder[1]|2:LIMit:UPPer <numeric_value>
This command sets the maximum scaling value for the specified recorder
output. The units used are dependent on the units currently set for the
CALC block specified in OUTPut:RECorder[1]|2:FEED <data_handle>.
Syntax
OUTP
Parameters
Example
OUTP:REC:LIM:UPP 10
Reset Condition
On reset, the maximum scaling value is set to +20 dBm.
N1911A/1912A P-Series Power Meters Programming Guide
1
:REC
2
Item
Description/Default
numeric_value
A numeric value for the minimum scaling
value. The units used—dBm, W or %—are
dependent on the units currently set for the
CALC block specified in
OUTPut:RECorder[1]|2:FEED
<data_handle>.
:UPP
:LIM
This command sets the maximum scaling
value to 10.
OUTput Subsystem
numeric_value
Space
?
Range of Values
–150 to +230 dBm
1 aW to 100 XW
0 % to 999 %
8
297