Error Messages - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
MEASurement Commands
Example
FETC2:POW:AC?

Error Messages

• If the last measurement is not valid error –230, "Data corrupt or stale"
• If the expected_value and resolution parameters are not the same as
112
Item
Description/Default
resolution
A numeric value for the resolution. If it is
unspecified the current resolution setting is
used. If a value is entered it should correspond
to the current resolution setting otherwise an
error occurs.
source list
The channel which the command is
implemented on. If unspecified the current
window setup is used. However, on the
N1912A, if the window shows a ratio or
difference measurement, the upper window
defaults to Channel A and the lower window
to Channel B.
1
The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the
command sub-systems. The parameters must be entered in the specified order. If parameters are
omitted, they default from the right. The parameter DEFault is used as a place holder. Specifying
DEF leaves the parameter value unchanged.
2
When the measurement result is linear this parameter represents the number of significant
digits. When the measurement result is logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001
respectively.
occurs. A measurement is valid after it has been initiated. It becomes
invalid when either a reset occurs or any measurement parameter, for
example frequency, is changed.
the current expected value and resolution setting on the specified
window, error –221, "Settings conflict" occurs.
This command queries the lower
window/upper measurement result.
N1911A/1912A P-Series Power Meters Programming Guide
Range of Values
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF
(@1)
(@2) (N1912A only)