[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STARt
<numeric_value>
This command sets the start frequency of average or peak frequency
sweep. It must be used in conjunction with external trigger.
If frequency sweep is disabled (frequency sweep step set to 0), start
frequency will be set but will not take effect.
This command is only applicable when used with E4410, N8480 (excluding Option CFT),
N O T E
E9300. E9320 or N1920 sensor.
SENS:FREQ:STAR, SENS:FREQ:STOP and SENS:FREQ:STEP are allowed to be
N O T E
set in any desirable sequence.
When frequency sweep mode is configured with frequency step size within its allowable
range, 1 to 2048, the following applies:
• If frequency stop point is greater than frequency start point, the frequency range will be
• If frequency stop point is less than frequency start point, the frequency range will be
• If frequency stop point and frequency start point are equal, it is the same as power
N1911A/1912A P-Series Power Meters Programming Guide
swept in an ascending order.
swept in a descending order.
sweep mode.