STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains information
on the calibrating status of the power meter.
The following bits in these registers are used by the power meter:
These bits are set at the beginning of zeroing (CALibration:ZERO:AUTO
ONCE) and at the beginning of calibration (CALibration:AUTO ONCE). Also
for the compound command/query CALibration[:ALL]?, this bit is set at
the beginning of the calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
N1911A/1912A P-Series Power Meters Programming Guide
Bit Number
Decimal
Weight
0
-
1
2
2
4
3-15
-
:OPER
STAT
Definition
Not used
Channel A CALibrating Status
Channel B CALibrating Status (N1912A only)
Not used
:SUMM
:CAL
STATus Subsystem
11
477