PPE
Once the power meter has received a PPC command, the PPE (Parallel Poll
Enable) secondary command configures the power meter to respond to a
parallel poll on a particular data line with a particular level.

Table 17-95PPE Mapping

PPU
The PPU (Parallel Poll Unconfigure) command disables the power meter
from responding to a parallel poll. This is effectively a universal disable.
760
Bit
Weight
Meaning
0
1
Bit positions for response:
000 (bit 0), 001 (bit 1), 010 (bit 2), 011 (bit 3),
1
2
100 (bit 4), 101 (bit 5), 110 (bit 6), 111 (bit 7)
2
4
3
8
Sense bit
0 - response bit is cleared during a parallel poll
if requesting service.
1 - response bit is set during a parallel poll if
requesting service.
4
16
Always 0
5
32
Always 1
6
64
Always 1
7
128
Always 0
N1911A/1912A P-Series Power Meters Programming Guide