[SENSe[1]]|SENSe2:SWEep[1]|2|3|4 Commands

These commands set offset time and time gate length as illustrated in the
following diagram:
Offset time and time gate length values can be set for up to four
measurement gates per channel. Measurement gate number is defined by
the numeric value following the SWEep component of the command.
These commands can only be used with P-Series and E9320 sensors. The E9320 sensor
N O T E
must be set to NORMal mode.
The following commands are detailed in this section:
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME <numeric_value>
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
432
Pre-
Trigger
With no
trace:
internally
calculated
time
Incoming signal
from sensor
Trigger
Point
Defined using:
TRIG:LEVel
TRIG:SLOPe
TRIG:HYSTeresis
N1911A/1912A P-Series Power Meters Programming Guide
With no trace: internally calculated time
Trigger
Delay
TRIG:DEL
Delayed
Trigger
SENSe:SWEep:OFFSet:TIME
Point
Data Collection Time
Post-Trigger
Time gate length:
SENSe:SWEep:TIME
Offset time: