Trigger[1]|2[:Immediate] - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

TRIGger[1]|2[:IMMediate]

This command causes a trigger to occur immediately, provided the
specified channel is in the wait for trigger state. When this command is
executed, the measurement result is stored in the power meter's memory.
Use FETCh? to place the measurement result in the output buffer.
TRIGger[1]|2:DELay:AUTO is ignored if TRIGger[1]|2[:IMMediate] is
set to ON.
This command performs the same function as INITiate[1]|2:[IMMediate].
N O T E
Syntax
Example
TRIG
Error Messages
If the power meter is not in the wait for trigger state, then
TRIGger:IMMediate causes error –211, "Trigger ignored".
N1911A/1912A P-Series Power Meters Programming Guide
TRIG
1
2
:IMM
This command causes a Channel A trigger
to occur immediately.
TRIGger Subsystem